AORM Software Package
923133 Rev A ISSUED: June 2013 91
The 3T timing jitter is calculated by taking the standard deviation of the difference between each
width and the 3T mean. This is 3.214 ns, normalized by:
Timj
T3
3214
100%
2315
1389%=⋅ =.
.
.
The 4T timing
jitter cannot be calculated because there is only one value (at least two values are
required).
The 5T timing jitter is +6.109%.
The overall timing jitter is calculated using a weighting formula, which results in the standard
deviation of the mean centered distributions. In this example, it is calculated as:
() ()
()
Timj
overall
=
⋅−+
+−
=
1389 3 1 6109 2 1
321
3208%
22
..
.