Teledyne-lecroy QPHY-LPDDR2 Uživatelský manuál Strana 22

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 49
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 21
22
QPHY-LPDDR2-OM-G Rev B
PCF200 Fixture Overview
Probe Connection to PCF200
Probe Calibration Menu
D620 Probe Calibration
Advanced mode is available:
Advanced Mode Probe Calibration Menu
Advanced Probe Calibration
Read
about
PCF200
fixture
Connect
PCF200
fixture to
oscillocope
Access
Probe
Calibration
Menu
Perform
Probe
Calibration
Connect
Probe to
Fixture
Disconnect
Probe from
Fixture
Connect Probe
to Circuit and
Make
Measurements
Repeat
this
procedure
for each
probe
Connecting probes to the circuit under test can be a difficult procedure. With
this in mind, LeCroy's system is de
signed in a manner that allows you to set up
the probe calibration fixture, calibrate each individual probe once, connect your
probes to the circuit, and disconnect the fixture. Once your probes are in the
circuit, there is no need to revisit the fixture until the next calibration interval.
You should familiarize yourself with the following topics:
PCF200 Fixture Overview
Probe Connection to PCF200
Probe Calibration Menu
D620 Probe Calibration
Advanced mode is available:
Advanced Mode Probe Calibration Menu
Advanced Probe Calibration
Zobrazit stránku 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 48 49

Komentáře k této Příručce

Žádné komentáře